DIATHERM M-1000

SOLID STATE ELCTROSURGICAL UNIT, MODEL M-1000

M1000 ELECTRO-SURGICAL UNIT (Diathermy) uses two industry standar Microcontrollers. One Microcontroller is dedicated for Display, Keyboard scan and some other user specific actions. The other Microcontroller is used for the overall control. These two microcontrollers communicate with each other to share data. The different waveforms Cut, Coagulation, etc) are generated in hardware from a Quartz Crystal controlled oscillator and are selected bythe~control Microcontroller based on the user selection. This waveform is then feed to a Raciio Frequency (RF) amplifier to generate the final RF output.

Feedback Control:

One of the most important functions of the Control Microcotroller is to provide feedback to regulate the output power of the Diathermy.

Salient Features :

1. Dual Microcontroller for generation of Output/dosage & other main functions

2. Dig!tal readout in all modes, user friendly features, control by soft touch key.

3. Delivers two cut.three coagulation & two bipolar modes.

4. Flat front panel design,easy access to controls & visibility of digital display.

5. Nine user selectable programme.save & recall mode.

6. Quartz controlled o/p wave frequency, active feedback by tissue resistance sense.

7. Completely Isolated output under water cutting unit.

8. Programmed self diagnostic alarm system.

9. Dosage error and time out sense.

Technical Specification:

A. Monopolar mode (at 300 ohms load)

1. Cut power–0 to 300 Watts .:!: 15%,

sine shaped wave, 400 Khz .:!: 1 %

Crest Factor =1.9.

2. Blend power-0 to 200 Watts 1 5%, burst of sine shaped wave at

55% Duty cycle, Crest factor = 3.8

3. Desiccate Coagulation mode 0 to 150 Watts 1 5% ,

Damped sine shaped wave at 10 micro sec. interval, Crest factor = 4.6

4. Fulgurate Coagulation mode 0 to 120 Watts 1 5% Damped Sine wave

at an interval of 20 micro sec. Crest factor= 6.3

5. Spray Coagulation-0 to 100 Watts :!:_1,5%D amped sine shaped wave

at 30 micro sec. interval, Crest factor= 7.5

B. Bipolar mode (at 100 ohms load)

1. Bipolar Coagulation-0 to 60 Watts _:!:15%,sines haped wave, 400 Khz, Crest Factor= 2.0.

2. Bipolar Cut-0 to 120 Watts _:!: 15%,sine shaped wave, 400 Khz, Crest Factor= 2.0.

C. Leakage Current

1. High frequency-Monopolar less than 150 mA & Bipolar less than 50mA rms.

2. AC mains at 200 to 240 Volts AC-less than 50 microAmp (Open ground)

POWER INPUT:

Voltage 200 240 Volts A.C, 50 Hz,

Maximum Power Consumption-600 Watts (Approx)

DIMENSIONS & WEIGHT:

Height = 185mm.(7.25inches) x Width= 375mm.(14.75inches) x

Depth= 300mm.(11.75 inches); Weight= 7.5 Kgs (Approx).

Warranty: