Model M-300 - Technical Specifications

A. Monopolar Mode (at 300 ohms load)

Cut Power 0 to 300 Watts ± 15%, sine shaped wave, 400 KHz ± 1%
Crest Factor = 1.9
Blend power 0 to 200 Watts ± 15%, burst of sine shaped wave at 55% duty cycle
Crest Factor = 3.8
Desiccate Coagulation 0 to 150 Watts ± 15%, Damped sine shaped wave at 10 micro sec. interval
Crest Factor = 4.6
Fulgurated Coagulation 0 to 120 Watts ± 15%, Damped Sine wave at 20 micro sec. interval
Crest Factor = 6.3
Spray Coagulation 0 to 100 Watts ± 15%, Damped sine shaped wave at 30 micro sec. interval
Crest Factor = 7.5

B. Bipolar Mode (at 100 ohms load)

Bipolar Coagulation 0 to 60 Watts ± 15%, sine-shaped wave at frequency of 400 KHz ± 1%
Crest Factor = 2.0
Bipolar Cut 0 to 120 Watts ± 15%, sine-shaped wave, 400 KHz
Crest Factor = 2.0

C. Leakage Current

High frequency - Monopolar Less than 150 mA & Bipolar less than 50 mA rms
A.C omens @ 200 to 240 Volts A.C-less than 50 microAmp (Open ground)

Dimensions

Size Height = 4" (12.7 cm) × Width = 14.9" (38 cm) × Depth = 12.9" (33 cm)
Weight 6.0 kg (Approximate, without accessories)

Power Input

Input Voltage 200 - 240 Volts A.C, 50 Hz
Maximum Power Consumption 600 Watts (Approximate)

Model M-600/M-1000 - Technical Specifications

Output Specifications - Monopolar Mode (at 300 ohms load)

Cut Power 0 to 150 Watts ± 15%, sine shaped wave, 400 KHz ± 1%
Crest Factor = 1.9
Blend power 0 to 150 Watts ± 15%, burst of sine shaped wave
Crest Factor = 3.2
Coagulation Power 0 to 150 Watts ± 15%, Damped sine shaped wave

Output Specifications - Bipolar Mode (at 100 ohms load)

Bipolar Coagulation 0 to 60 Watts ± 15%, sine-shaped wave, 400 KHz, Crest Factor = 2.0
Bipolar Cut 0 to 120 Watts ± 15%, sine wave, 400 KHz, Crest Factor = 2.0

Safety & Construction

Leakage Current - Monopolar Less than 150 mA rms
Leakage Current - Bipolar Less than 50 mA rms
Power Input Input Voltage 200 - 240 Volts A.C, 50 Hz
Maximum Power Consumption 600 Watts (Approximate)
Size Height = 185mm (7.25 inches) × Width = 375mm (14.75 inches) × Depth = 300mm (11.75 inches)
Weight 7.5 Kgs (Approximate)

General Features & Certifications

Quality Standards

  • Manufactured using ISO standard components
  • Meets international safety requirements
  • Complete electrical isolation from mains
  • Low RF leakage current as per international standards

Safety Features

  • Patient plate monitoring with audio & visual indicators
  • Automatic power regulation with feedback control
  • Self-diagnostic alarm system
  • Over-temperature protection
  • Error detection and timeout sensing

User Interface

  • Digital display readout in all modes
  • User-friendly controls
  • Soft touch keypad (M-1000)
  • Footswitch and handswitch operation
  • Nine programmable modes (M-1000)

Warranty

  • One year from date of purchase
  • Comprehensive after-sales support
  • Spare parts availability

Have Technical Questions?

Our expert team is available to assist with detailed specifications and applications